top of page

Easy Blood Gas Technical Specifications 

Sample Type: Whole blood

 

Sample Size: 100 uL Syringe 75 uL Capillary

​

Method: Direct measurement by Ion Selective Electrode (ISE)

 

Measured Parameters

 

pH : 6.900 – 7.900 pH units

PCO2 : 8.0 – 150.0 mmHg

PO2 : 10 – 700 mmHg

 

 

Calculated Parameters

 

pH (T) : pH temperature corrected

PCO2 (T) : PCO2 temperature corrected

PO2 (T) : PO2 temperature corrected

TCO2 (Total Carbon Dioxide) :  0 – 50 mmol/L

HCO3 (Bicarbonate) :   0 – 50 mmol/L

BEb (Base Excess in blood) :  -25.0 – 25.0 mmol/L

BEecf (Base Excess in extra cellular fluid) :  -25.0 – 25.0 mmol/L

SBC (Standard Bicarbonate) :  0 – 50 mmol/L

%SO2C (Oxygen Saturation calculated at normal P50) :  40.0 – 100.0%

A-aDO2 (Alveolar arterial oxygen gradient) :  0 – 700 mmHg

RI (Respiratory Index) :  0.0 – 70.0

 

Input Parameters

 

Patient Temperature :  5 – 45°C

Haemoglobin :  30 – 30.0 g/dL

FIO2 :  10 – 100%

Patient ID :  14 digits

 

Sample Temperature Control: 37°C +/– 0.2°C

 

Ambient Conditions: 15 – 30°C (59 – 86°F), 500 – 800 mmHg (max 15 PSI)

5 – 85% relative humidity, non-condensing atmospheric air environment (21% PO2)

 

Analysis Time:  125 seconds

 

Data Storage: 64 Patient results with Operator ID, Patient ID, Date and Time

QC – up to 30 results for each Level (1, 2, 3)

 

Calibration: Automatic or On-Demand

 

Input/Output: Numeric keypad, graphic display, 27 column thermal printer,

barcode reader port, RS-232 computer interface port

 

Power: 100/115 ~ VAC, 50 – 60 Hz, 0.8 A or 220 ~ VAC, 50 – 60 Hz, 0.4A

 

Size & Weight: 14.5" W x 12.5" H x 7" D (37 cm W x 32 cm H x 18 cm D)

16 lbs. (7.3 kg) with reagent module

bottom of page