Easy Blood Gas Technical Specifications
Sample Type: Whole blood
Sample Size: 100 uL Syringe 75 uL Capillary
​
Method: Direct measurement by Ion Selective Electrode (ISE)
Measured Parameters
pH : 6.900 – 7.900 pH units
PCO2 : 8.0 – 150.0 mmHg
PO2 : 10 – 700 mmHg
Calculated Parameters
pH (T) : pH temperature corrected
PCO2 (T) : PCO2 temperature corrected
PO2 (T) : PO2 temperature corrected
TCO2 (Total Carbon Dioxide) : 0 – 50 mmol/L
HCO3 (Bicarbonate) : 0 – 50 mmol/L
BEb (Base Excess in blood) : -25.0 – 25.0 mmol/L
BEecf (Base Excess in extra cellular fluid) : -25.0 – 25.0 mmol/L
SBC (Standard Bicarbonate) : 0 – 50 mmol/L
%SO2C (Oxygen Saturation calculated at normal P50) : 40.0 – 100.0%
A-aDO2 (Alveolar arterial oxygen gradient) : 0 – 700 mmHg
RI (Respiratory Index) : 0.0 – 70.0
Input Parameters
Patient Temperature : 5 – 45°C
Haemoglobin : 30 – 30.0 g/dL
FIO2 : 10 – 100%
Patient ID : 14 digits
Sample Temperature Control: 37°C +/– 0.2°C
Ambient Conditions: 15 – 30°C (59 – 86°F), 500 – 800 mmHg (max 15 PSI)
5 – 85% relative humidity, non-condensing atmospheric air environment (21% PO2)
Analysis Time: 125 seconds
Data Storage: 64 Patient results with Operator ID, Patient ID, Date and Time
QC – up to 30 results for each Level (1, 2, 3)
Calibration: Automatic or On-Demand
Input/Output: Numeric keypad, graphic display, 27 column thermal printer,
barcode reader port, RS-232 computer interface port
Power: 100/115 ~ VAC, 50 – 60 Hz, 0.8 A or 220 ~ VAC, 50 – 60 Hz, 0.4A
Size & Weight: 14.5" W x 12.5" H x 7" D (37 cm W x 32 cm H x 18 cm D)
16 lbs. (7.3 kg) with reagent module