Olympus AU 640 Clinical Chemistry Analyzer

  • Manufactured By Olympus Diagnostics
  • Fully automated
  • Throughput 800 test / hour 1200 with ISE
  • Fully open system
  • Low reagent volume 150 – 550 µl
  • 99 parameters
  • Spectrophotometer

DescriptionTechnical Specs

Fully automated analyzer, for routine, STAT, urine and homogeneous immunoassays.

Analytical methods available are Clinical chemistry and immunological parameters. End point, kinetic assays, fixed-time-kinetics, optionally ISE

Simultaneously processed 48 photometric tests, 51 with ISE

Throughput of 800 photometric tests/hour; maximum of 1,200 with ISE

Sample feeder Racks with 10 samples each (bar codes on primary tubes and on racks); capacity of 150 samples; continuous loading

Sample tubes In primary and secondary tubes; diameter between 11.5 and 16.5 mm; height between 55 and 102 mm

Up to 22 positions for STAT samples, bar coded primary tubes

Sample Volume 2 – 50 µl in 1 µl steps (1 – 50 µl for repeats)

Reagent Supply 48 positions for R1 and R2; refrigerated (4 °C – 12 °C)

Reagent Volume 1. reagent: 25 – 300 µl; 2. reagent: 25 – 300 µl; (in 1 µl steps)

Reaction volume 150 – 550 µl

Reaction Cuvettes made of Quartz

Reaction time Up to 8 minutes, 40 seconds

H x W x D 61 x 149 x 59 (cm)
24 x 59 x 23 (in)
Weight 160kg (350 lb)
Thermal Control 37C +/- 0.5 C
Plumbing Self-contained
Voltage Requirements 100 – 120 V 50/60 Hz
220 – 240 V 50/60 Hz
Environment Ambient operating temperature from 18 – 30C Humidity 20 – 80% (noncondensing)
Microprocessor (CPU) Multiple, high integration, 16-bit microprocessors 40 MB hard disk drive
3.5” double-sided, high-density,
1.44 MB floppy diskette drive facilities software updates and archiving data

Copyright © 2020 Tek Medical. Northway House, 1379 High Road, London N20 9LP. United Kingdom