Medica Easy RA - Veterinary Clinical Chemistry Analyzer

  • Easy to afford
  • Easy to use, learn and maintain.
  • Virtually maintenance free
  • Up to 340 test / hour
  • 24 reagents position with on board cooling
  • 24 sample positions with provision for sample cups, or primary tubes
  • Zeno flash lamp, 5 years typical life

 

System specifications: true random access, clinical chemistry analyzer

 Throughput:  analysis rate, photometric: 150 tests/hour; Patient results per hour, photometric: up to 120 tests/hour. Patient results per hour, photometric with ISE: up to 340 tests/hour; STAT time: 8 minutes (Na+/K+/Cl-/CO2/GLU/BUN/CREA)

 Types of analysis:  endpoint, enzymatic, rate, bichromatic, potentiometric.

 Samples : 24 sample positions per sample ring for patient, calibrator, or QC samples; STAT: up to 5 user-defined positions; optional second sample ring that is uniquely identified by analyzer; automatic dilution: 1:1 and 1:2

 Sample volume : photometric chemistries: 2.0–25.0 μL; programmed in 0.1 μL steps; ISE chemistries: serum: 80–90 μL; urine: 140 μL

 Sample containers: sample cups or primary tubes in a wide range of sizes,

 Sample identification: position ID, barcode ID (optional), barcode types: codabar, code 39, 128, interleaved 2 of 5

Reagents: 24 positions for reagents; reagent cooling temperature 12o–15o less than ambient (with refrigeration option); reagent identification: RFID (radio frequency identification) technology—automatic tracking and entry of reagent information (chemistry name, lot number, expiration date; reagent volumes; analysis volumes for reagents, samples, diluent; primary and secondary wavelengths; reaction read times; analysis type; reagent and sample blanking; linear range of assay; acceptable absorbance ranges). Reagents are ready to use.

 Reagent volumes: reagent volume (R1)/test 120–350 μL; programmed in 1 μL steps. reagent volume (R2)/test 10–50 μL; programmed in 1 μL steps.

 Water supply: reagent grade deionized water, diluent bottle

 Sampling system:  probe pre-heater; single probe with RF level sensing; inner and outer probe washing

 Cuvette material: optical acrylic; disposable segments; 12 cuvettes per segment; 6 total segments in reaction area

 Reaction time: 1–15 minutes

Reaction temperature:  37o C  +/- 0.25o C (photometric chemistries)

 Wavelengths: 340, 405, 520, 550, 600, 700; half bandwidth 10 +/- 2 nm

Light source: xenon flash lamp, 5 year typical life

 Photometric linearity:  0.0–2.0 Abs units for 1 cm pathlength (1% deviation)

 Photometric resolution:  0.0001 Abs units at 1.0 Abs

Quality Control: 2 levels of controls (Levey-Jennings plots for two levels)

 Calibration curves:  single and multilevel calibration (based on analyte).

 User interface:  edit and monitor worklists; review results; review calibration and quality control results; Levey-Jennings charts for 31 days of QC results; on-board diagnostics and individual component monitoring; graphic instructions for daily, weekly and monthly maintenance procedures

 Data storage : 2000 patient results; 56,000 test results

 Power requirements: 100 VAC–240 VAC } 10% 50–60 Hz, 4.0/2.0A

 Size and weight:  40˝ w x 15˝ h x 26˝ d (102 cm x 38 cm x 66 cm), 88 lbs (40 kgs) without reagents

Ambient conditions: 15 o C – 30o C (59o–86o F); <85% relative humidity, non-condensing atmospheric air environment

Computer requirements: minimum: Windows XPR; CD/CD-RW; 1 RS-232 or USB port; touch screen monitor or SVGA color monitor, mouse and keyboard printing local or network printer

 Optional features: ISE Module; refrigeration; laboratory information system; (LIS) connectivity option (ASTM protocol);

on-board sample bar code reader

 

 

   