Horiba ABX Pentra 60

  • 5 Part Differential
  • Highly compact and cost-effective
  • 60 samples/hour
  • Micro-sampling (CBC: 30 µL - DIFF: 53 µL)
  • Cytochemistry, Impedance (real cell volume measurement)
  • Optical internal cellular structure by light absorbance
Perfect homogenization of blood samples with reagents - MDSS** Technology
Reagents: Only 4 onboard reagents and 1 diluent
Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology
3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix
Basophils counted through specific channel
High resolution matrix includes the determination of 2 additional subpopulations (% and #): Atypical Lymphocytes (ALY***) and Large Immature Cells (LIC***)    
Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS* Technology
 
26 Parameters : 
WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, ALY# & ALY%*, LIC# & LIC%*, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT*, PDW*.
Blood Volume : 
CBC mode : 30 µL
CBC + DIFF mode : 53 µL
 
Specifications:
 
WBC
Precision (%CV) < 2.0
 
Linearity limits 0-120 x 10³/µL
 
RBC
 
Precision (%CV) < 2.0

Detailed description here

Technical Spec Value

Linearity limits 0-8 x 106/µL
 
HGB
 
Precision (%CV) < 1.0
 
Linearity limits 0-24 g/dl
 
HCT
 
Precision (%CV) < 2.0
 
Linearity limits 0-67%

PLT (whole blood)
 
Precision (%CV) < 5.0
 
Linearity limits 0-1900 x 10³/µL
 
PLT (PLT concentrate)
 
Precision (%CV) < 5.0
 
Linearity limits 0-2800 x 10³/µL